The study was performed to compare the efficacy and safety during radiofrequency ablation (RFA) using ThermoCool SmartTouch (ST) and ThermoCool SmartTouch-SF (STSF) catheters in the porcine heart. RFA was performed on the porcine myocardium by using two irrigated ablation catheters. Three groups were divided based on the different contact forces (CFs) low contact force (LCF) (1-3 g), medium contact force (MCF) (5-10 g), and high contact force (HCF) (15-20 g). In each group, RFA was delivered at four power settings of 30, 40, 50, 60 W. At