Secondary neutron dose, as estimated by Monte Carlo simulations, was found to significantly increase the odds of CIED resets by 55% per mSv. Clinically significant battery depletion was observed in 5 devices. We observed no noise, over- or undersense, pace inhibition, or inappropriate shock therapy during 362 fractions of live monitoring. Reprogrammable CIED reset was the most commonly observed malfunction during proton therapy, and reset risk depended on secondary neutron exposure. The benefits of proton therapy are expected to outweigh the